FDA Adverse Event Injury Summary report: N

M2A RINGLOC ACETABULAR LINER 28MM SIZE 25 STANDARD

MDR report key: 2060176 · Received April 15, 2011

Report

Report Number
1825034-2011-00271
Event Type
Injury
Date Received
April 15, 2011
Date of Event
March 16, 2011
Report Date
March 22, 2011
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
K002379
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. DATE IMPLANTED - UNKNOWN. EVALUATION OF THE RETURNED COMPONENT FOUND EVIDENCE OF DAMAGE FROM SURGICAL REMOVAL. THE ORIGINAL MACHINE LINES WERE PRESENT ON THE BACKSIDE OF THE POLYETHYLENE LINER EXCEPT FOR A NARROW BAND CLOSE TO THE APEX. THE LINER SHOWED CIRCULAR WITNESS MARKS FROM CONTACT WITH THE SCREW HOLES. THE ARTICULAR SURFACE OF THE COCR INSERT SHOWED EVIDENCE OF LIGHT SCRATCHING. THERE WAS NO EVIDENCE OF CHANGES IN SURFACE CONTOUR AT THE EDGES OF THE CONTACT AREA. THE OUTER RIM OF THE INSERT APPEARED TO SHOW EVIDENCE OF MINOR DEFORMATION IN ONE LOCATION. BASED UPON THE APPEARANCE OF THE COMPONENT, WEAR RATES DID NOT APPEAR TO BE EXCESSIVE. THE USER FACILITY WAS NOTIFIED OF THE EVENT ON (B)(6), 2011. TO DATE, A RESPONSE HAS NOT BEEN RECEIVED FROM THE USER FACILITY. IN THE EVENT THAT THE USER FACILITY SUBMITS A MEDWATCH REPORT, BIOMET WILL FORWARD A COPY TO THE FDA. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT TOTAL HIP ARTHROPLASTY ON UNKNOWN DATE. SUBSEQUENTLY, THE PATIENT WAS REVISED ON (B)(6), 2011 DUE TO PATIENT CONCERN OF METAL IONS. THE SURGEON NOTED DURING THE REVISION THAT THE COMPONENTS WERE WELL FIXED AND THERE WAS NO SIGN OF METAL DEBRIS. THE ACETABULAR LINER AND MODULAR HEAD WERE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M2A RINGLOC ACETABULAR LINER 28MM SIZE 25 STANDARD PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 824980

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R