FDA Adverse Event Malfunction Summary report: N

DISCOVERY II

MDR report key: 2060132 · Received April 15, 2011

Report

Report Number
2124215-2011-05781
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
July 29, 2010
Report Date
October 26, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME THE PRODUCT REMAINS IN SERVICE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WILL BE UPDATED AS NECESSARY.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE BATTERY STATUS HAS BEEN FLUCTUATING MAKING IT DIFFICULT TO PREDICT LONGEVITY REMAINING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DISCOVERY II IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1283

Patients

Seq Age Sex Outcome Treatment
1 1283