FDA Adverse Event Malfunction Summary report: N

VITALITY

MDR report key: 2060062 · Received April 15, 2011

Report

Report Number
2124215-2011-05607
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
March 29, 2011
Report Date
April 4, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS REPORT WILL BE UPDATED WHEN ANALYSIS IS COMPLETE.

Additional Manufacturer Narrative · 1

A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE DEVICE DECLARED END OF LIFE (EOL) POST EXPLANT AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI AND EOL WERE TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED DUE TO NORMAL BATTERY DEPLETION. INITIAL LABORATORY ANALYSIS FOUND THE DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T125

Patients

Seq Age Sex Outcome Treatment
1 62 YR T125| 4136| E110| 0158