FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2060054 · Received April 15, 2011

Report

Report Number
2124215-2011-06043
Event Type
Injury
Date Received
April 15, 2011
Date of Event
March 30, 2011
Report Date
March 30, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
PMA / PMN Number
P960004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE AVAILABLE INFORMATION SUGGESTS THIS DEVICE AND LEAD REMAIN IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER AND RIGHT VENTRICULAR PACING LEAD EXHIBITED INCREASED THRESHOLD MEASUREMENTS. AUTOMATIC CAPTURE HAD BEEN PROGRAMMED ON AT A PREVIOUS DEVICE CHECK. UPON INTERROGATION, THE DEVICE WAS FOUND TO BE IN RETRY AT 3.5V AT .4 MS. A STORED EPISODE REVEALED LOSS OF CAPTURED WITH ASYSTOLE FOR 2.1 SECONDS. THE PATIENT HAS BEEN EXPERIENCING SHORTNESS OF BREATH AND FATIGUE. OUTPUTS WERE REPROGRAMMED TO 6.5 V AT 1.0 MS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4470

Patients

Seq Age Sex Outcome Treatment
1 87 YR Life Threatening S601| 4470