FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 2060043 · Received April 15, 2011

Report

Report Number
2124215-2011-05705
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
March 18, 2011
Report Date
June 9, 2017
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS OF TODAY, NO ADDITIONAL INFORMATION IS AVAILABLE AND AT THIS TIME, OUR INVESTIGATION IS COMPLETE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

SUBSEQUENT INFORMATION INDICATES THAT THE DEVICE WAS EXPLANTED AFTER THE PATIENT PASSED AWAY. THE DEVICE WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS CARDIAC RESYNCHRONIZATION THERAPY-DEFIBRILLATOR (CRT-D) DISPLAYED AN ATRIAL ARRHYTHMIA THAT BEGAN IN THE MONITOR ONLY ZONE AND MET DURATION. THE DEVICE WENT IN TO RE-DETECTION AND ALL DETECTION ENHANCEMENTS WERE DISABLED. AT SOME POINT, THE RHYTHM ACCELERATED INTO A HIGHER ZONE AND THE PATIENT THEREFORE RECEIVED ONE 41 JOULE SHOCK. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT. THE DEVICE REMAINS IN SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COGNIS IMPLANTABLE CHF PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 68 YR 4136| 4549| 4543| N119| 0184