FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.5CCM 30GA 8MM BLS 100BX/500 AP

MDR report key: 20599875 · Received November 4, 2024

Report

Report Number
3024508819-2024-00424
Event Type
Malfunction
Date Received
November 4, 2024
Date of Event
October 15, 2024
Report Date
January 13, 2025
Manufacturer
EMBECTA MEDICAL I LLC - HOLDREGE, NE
Product Code
FMF
PMA / PMN Number
UNKNOWN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTIONS MADE TO SECTION H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, & INVESTIGATION CONCLUSIONS). INVESTIGATION SUMMARY: NO PHYSICAL SAMPLES WERE RECEIVED HOWEVER THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTO(S) PROVIDED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED, AND NO NON-CONFORMANCE'S WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. CAPA: PR00034 WAS OPENED IN JULY 2024 TO ADDRESS ILLEGIBLE/PARTIAL LABELING ON PACKAGING.

Additional Manufacturer Narrative · 0

5 UNITS IMPACTED FROM LOT # 3142108. SEE SECTION C FOR ADDITIONAL INFORMATION.

Description of Event or Problem · 0

THE NEEDLE IS COVERED BY CAP AND THE GLUE IS COLORLESS, IT IS VERY HARD FOR WORKERS TO FIND THE ISSUE AND PICK IT OUT DURING MASS PRODUCTION. NOTE: AS PER ATTACHMENT COUNTRY EVENT MENTIONED AS AU. BUT EMAIL RECEIVED FROM MALAYSIA. AWARENESS DATE AS 09-SEP-2024 BUT EMAIL RECEIVED ON 15-OCT-2024. 18OCTOBER2024: PLEASE BE INFORMED THAT THE ARRIVAL STOCKS IN THE TABLE BELOW WERE FOUND 5 EA WITHOUT BATCH AND EXPIRY DATE ON PHYSICAL STOCK DURING REDRESSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
789225 SYRINGE 0.5CCM 30GA 8MM BLS 100BX/500 AP SYRINGE, PISTON FMF EMBECTA MEDICAL I LLC - HOLDREGE, NE 326725 3142108

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown