FDA Adverse Event
Malfunction
Summary report: N
LP6+ VOLUME VENTILATOR
MDR report key: 205969
·
Received January 6, 1999
Report
- Report Number
- 2183157-1999-00005
- Event Type
- Malfunction
- Date Received
- January 6, 1999
- Date of Event
- December 11, 1998
- Report Date
- January 6, 1999
- Manufacturer
- NELLCOR PURITAN BENNETT
- Product Code
- CBK
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
DURING SVC OF UNIT A STOP CYCLE WITH ALL LEDS AND CONSTANT SINGLE TONE ALARM WAS FOUND, DUE TO INTEGRATED CIRCUIT U28 ON THE LOGIC BOARD BEING OUT OF SPECIFICATION. REPLACED U28.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LP6+ VOLUME VENTILATOR | VOLUME VENTILATOR | CBK | NELLCOR PURITAN BENNETT | LP6+ | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |