FDA Adverse Event Malfunction Summary report: N

LP6+ VOLUME VENTILATOR

MDR report key: 205969 · Received January 6, 1999

Report

Report Number
2183157-1999-00005
Event Type
Malfunction
Date Received
January 6, 1999
Date of Event
December 11, 1998
Report Date
January 6, 1999
Manufacturer
NELLCOR PURITAN BENNETT
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING SVC OF UNIT A STOP CYCLE WITH ALL LEDS AND CONSTANT SINGLE TONE ALARM WAS FOUND, DUE TO INTEGRATED CIRCUIT U28 ON THE LOGIC BOARD BEING OUT OF SPECIFICATION. REPLACED U28.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LP6+ VOLUME VENTILATOR VOLUME VENTILATOR CBK NELLCOR PURITAN BENNETT LP6+ NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other