GLENOSPHERE ORIENTATION GUIDE
Report
- Report Number
- 1818910-2024-22940
- Event Type
- Malfunction
- Date Received
- November 1, 2024
- Date of Event
- October 23, 2024
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- PXH
- UDI-DI
- 10603295116592
- PMA / PMN Number
- K170748
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE.
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY AN ANALYSIS OF THE PRODUCT COULD NOT BE PERFORMED SINCE A PHYSICAL SAMPLE WAS NOT RECEIVED FOR EVALUATION. AN EVALUATION OF THE MANUFACTURING RECORD COULD NOT BE PERFORMED AS THE REQUIRED PRODUCT IDENTIFICATION NUMBER WAS NOT PROVIDED TO COMPLETE THE EVALUATION. AS PART OF OUR COMPANY QUALITY SYSTEM PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND DISTRIBUTED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF THE POST-MARKET SURVEILLANCE. HOWEVER, IF THE PRODUCT OR PRODUCT ID NUMBERS ARE RECEIVED AT A LATER DATE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE.
IT WAS REPORTED THAT BOTH XTEND GLENOSPHERE ECCENTRICITY HANDLES WERE STARTING TO CRACK AT THE GLENOSPHERE INTERFACE WHILE DIALING IN ECCENTRICITY OF THE FINAL IMPLANT. SURGEON WAS ABLE TO SUCCESSFULLY DIAL ECCENTRICITY. ALL PIECES RETRIEVED FROM PATIENT. NO DELAY. NO LOOSENING. NO FAILURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 791813 | GLENOSPHERE ORIENTATION GUIDE | SHOULDER PROSTHESIS, REVERSE CONFIGURATION | PXH | DEPUY ORTHOPAEDICS INC US | 10603295116592 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Male |