FDA Adverse Event Malfunction Summary report: N

AUTOMATED IMPELLA CONTROLLER

MDR report key: 20578433 · Received October 31, 2024

Report

Report Number
1220648-2024-21651
Event Type
Malfunction
Date Received
October 31, 2024
Date of Event
May 26, 2021
Report Date
October 31, 2024
Manufacturer
ABIOMED, INC.
Product Code
OZD
UDI-DI
00813502010022
PMA / PMN Number
P140003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED AND EVALUATED, AND THE INVESTIGATION FOR THE REPORTED CONTROLLER ERROR (YELLOW ALARM) HAS BEEN COMPLETED. CONSOLE LOGS FROM THE REPORTED DAY OF EVENT ARE CONSISTENT WITH COMPLAINT AND SHOW A SINGLE OCCURRENCE OF KBD COMMUNICATION ERROR FOLLOWED BY ALARM #24 (LOSS OF COMMUNICATION WITH KBD). CONSOLE WAS POWERED ON AGAIN, THE NEXT DAY, BUT THE ALARM DID NOT REOCCUR. AFTER CONSOLE WAS BROUGHT TO (B)(6), THE ISSUE WAS INITIALLY NOT REPRODUCED DURING SEVERAL POWER-CYCLES. THE KBD ASSEMBLY AND ITS RIBBON CABLE WERE VISUALLY INSPECTED, AND NO OBVIOUS SIGNS OF DAMAGE OR DEGRADATION (E.G. CORROSION) WERE OBSERVED. IN THE COURSE OF TROUBLESHOOTING THE CONSOLE WAS SET UP TO POWER-CYCLE 500 TIMES (USING POWER-CYCLING FIXTURE), AFTER WHICH CONSOLE LOGS WERE ANALYZED. ANALYSIS REVEALED 13 OCCURRENCES OF KBD COMMUNICATION ERRORS AND 15 OCCURRENCES OF ALARM #24. AFTER THE 4-IN-1 WAS REPLACED AND TEST REPEATED, OCCURRENCES OF KBD-RELATED ERRORS AND ALARMS WAS REDUCED. THE CAUSE OF THE CONTROLLER ALARM UPON BOOT-UP WAS LOSS OF COMMUNICATION BETWEEN EPC AND KBD, MOST LIKELY DUE TO KBD ASSEMBLY MALFUNCTION. THIS REPORT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF HISTORICAL RECORDS.

Description of Event or Problem · 0

THE USER FACILITY'S BIOMEDICAL ENGINEER REPORTED THE AUTOMATED IMPELLA CONTROLLER (AIC) EXPERIENCED A CONTROLLER ERROR YELLOW ALARM. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2039717 AUTOMATED IMPELLA CONTROLLER CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS OZD ABIOMED, INC. IMPELLA CONTROLLER, PACKAGED, US 1125336 00813502010022

Patients

Seq Age Sex Outcome Treatment
1 0 DA Unknown