FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® EDTA 2K

MDR report key: 20559873 · Received October 29, 2024

Report

Report Number
1917413-2024-01010
Event Type
Malfunction
Date Received
October 29, 2024
Date of Event
October 3, 2024
Report Date
November 20, 2024
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
00382903678464
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED 1 USED SAMPLE AND 2 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR MISSING FILL LINE WAS OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLE WAS EVALUATED BY VISUAL EXAMINATION AND THE ISSUE OF MISSING FILL LINE WAS OBSERVED. 100 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED AND ALL MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE MISSING FILL LINE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THE ACTUAL DATE OF EVENT IS UNKNOWN. E1: INITIAL REPORTER FACILITY NAME: (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® EDTA 2K A TUBE WAS MISSING THE FILL LINE MARK. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® EDTA 2K A TUBE WAS MISSING THE FILL LINE MARK. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338726 BD VACUTAINER® EDTA 2K BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 4045118 00382903678464

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown