I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2024-00201
- Event Type
- Malfunction
- Date Received
- October 29, 2024
- Date of Event
- October 20, 2024
- Report Date
- December 3, 2024
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JPI
- PMA / PMN Number
- K183680
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 25-NOV-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING FOR LOT H24172 MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV AN, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CHEM8+ CARTRIDGE LOT H24172.
APOC INCIDENT # 964929 APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 25-OCT-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT HEMATOCRIT RESULT ON PATIENT. THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED NA CREA SAMPLE I-STAT 20-OCT-2024 09:57 09:57 124 MEQ/L 8.2 MG/DL A I-STAT 20-OCT-2024 11:04 11:04 124 MEQ/L 8.0 MG/DL B VS2 20-OCT-2024 NI 11:09 148 MMOL/L 4.1 MG/DL NI . AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 319074 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JPI | ABBOTT POINT OF CARE | NA | H24172 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |