FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 20557368 · Received October 29, 2024

Report

Report Number
2245578-2024-00201
Event Type
Malfunction
Date Received
October 29, 2024
Date of Event
October 20, 2024
Report Date
December 3, 2024
Manufacturer
ABBOTT POINT OF CARE
Product Code
JPI
PMA / PMN Number
K183680
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 25-NOV-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING FOR LOT H24172 MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV AN, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CHEM8+ CARTRIDGE LOT H24172.

Additional Manufacturer Narrative · 0

APOC INCIDENT # 964929 APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON 25-OCT-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT HEMATOCRIT RESULT ON PATIENT. THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED NA CREA SAMPLE I-STAT 20-OCT-2024 09:57 09:57 124 MEQ/L 8.2 MG/DL A I-STAT 20-OCT-2024 11:04 11:04 124 MEQ/L 8.0 MG/DL B VS2 20-OCT-2024 NI 11:09 148 MMOL/L 4.1 MG/DL NI . AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
319074 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JPI ABBOTT POINT OF CARE NA H24172

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown