I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2024-00197
- Event Type
- Malfunction
- Date Received
- October 29, 2024
- Date of Event
- October 5, 2024
- Report Date
- November 19, 2024
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- PMA / PMN Number
- K031739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 18-NOV-2024. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AN (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT S24124A.
ON (B)(6) 2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DESCREPANT RESULTS ON A 77 YEAR OLD MALE WITH PNEUMONIA WITH A PRODUCTIVE COUGH, AND CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULT(UG/L) SAMPLE: I-STAT, (B)(6) 2024, 14:30, 14:35, 0.14, WB. I-STAT, (B)(6) 2024, 14:30, 14:47, 0.06, WB. I-STAT, (B)(6) 2024, 14:30, 15:00, 0.03, WB. HS-TNT, (B))(6) 2024, 14:30, 17:45, 124, PLASMA. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THERE IS NO EVIDENCE THE PATEINT SUFFERED AN MI. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 715595-00V. REPORTABLE RANGE: 0.00 - 50.00. REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS). REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 916363 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | S24124 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Male |