ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2011-04850
- Event Type
- Injury
- Date Received
- April 14, 2011
- Date of Event
- March 10, 2011
- Report Date
- March 10, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
AS OF TODAY, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. IT IS SUSPECTED THAT THE DEVICE WAS BURIED WITH THE PATIENT. IF THE DEVICE IS RETURNED THE EVENT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT ONE DAY POST IMPLANT THIS CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND RIGHT VENTRICULAR (RV) DEFIBRILLATION LEAD WERE EXHIBITING HIGH OUT OF RANGE SHOCKING IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS. A PROCEDURE TOOK PLACE AND THE CONNECTIONS WERE VERIFIED ENSURING THE TERMINAL PIN WAS PAST THE CONNECTOR BLOCK. ONE DAY LATER, THE OUT OF RANGE MEASUREMENTS REAPPEARED. AFTER FURTHER EVALUATION, IT WAS COMMUNICATED THIS PATIENT HAD A CORONARY ARTERY BYPASS GRAFT PROCEDURE FIVE DAYS PRIOR TO THE CRT-D IMPLANT. THE PATIENT WAS ALSO VERY SWOLLEN DUE TO AIR IN THEIR CHEST FROM THE CHEST TUBE BEING REMOVED TOO SOON. IT WAS CONCLUDED THAT THE AIR IN THE PATIENT'S CHEST WAS HAVING AN AFFECT ON THE IMPEDANCE MEASUREMENTS. A DISK EVALUATION WAS SENT IN FOR ANALYSIS. FOUR DAILY SHOCK LEAD IMPEDANCE MEASUREMENTS RECORDED WERE 241, 101, 84, AND 60 OHMS. THE MOST RECENT MANUAL SHOCK LEAD IMPEDANCE VALUE WAS 59 OHMS. NO FURTHER ACTION PLANNED TO BE TAKEN AS THE PATIENT'S CONDITION AFFECTING THE SHOCK LEAD IMPEDANCE HAD RESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0180 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Life Threatening| R | 0180| N119| 4965| 4469 |