FDA Adverse Event Malfunction Summary report: N

IPC® FOOTPEDAL

MDR report key: 20554913 · Received October 29, 2024

Report

Report Number
1045254-2024-01887
Event Type
Malfunction
Date Received
October 29, 2024
Date of Event
August 29, 2024
Report Date
February 11, 2025
Manufacturer
MEDTRONIC XOMED INC.
Product Code
ERL
UDI-DI
00643169931398
PMA / PMN Number
K081277
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 1 899200, SERIAL/LOT #:(B)(6), UDI#: (B)(4). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

PRODUCT ANALYSIS FOR M5-09365 FOUND THAT NO FAULT FOUND AND DEVICE WORKING AS INTENDED. IMDRF CODE FOR M5-09365 IS A071801, B01, C19, D1102, F26 AND G04063. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS EVENT AND EVENT REPORTED UNDER REGULATORY REP #: 1045254-2024-01631 ARE FOR THE SAME EVENT. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS FOUND THAT HOUSING CRACKED. H6: ADDITIONAL INFORMATION SUGGEST THAT B17 , C20, D16 AND G0408201NO LONGER APPLY TO THIS EVENT. THE CORRECT AWARE DATE OF THE INITIAL REPORT WAS 4TH SEP-2024 MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT INTRA OP, DEVICE STARTED TO RUN ON OWN AFTER 20MIN OF BEING ATTACHED TO IPC POWER CONSOLE. IPC FOOT PEDAL ATTACHED BUT NOT ENGAGED BY ANYONE/ANYTHING. NO CONSEQUENCES OR IMPACT TO PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT INTRA OP, DEVICE STARTED TO RUN ON OWN AFTER 20MIN OF BEING ATTACHED TO IPC POWER CONSOLE. IPC FOOT PEDAL ATTACHED BUT NOT ENGAGED BY ANYONE/ANYTHING. NO CONSEQUENCES OR IMPACT TO PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303314 IPC® FOOTPEDAL DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) I ERL MEDTRONIC XOMED INC. 1898430 223394208 00643169931398

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11...."