ADULT NASAL CANNULA
Report
- Report Number
- 9611451-2011-00232
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- March 16, 2011
- Report Date
- March 17, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- CAT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT OPT546 ADULT NASAL CANNULAS ARE EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICES AND HAVE COMPLETED OUR INVESTIGATION.
(B)(4). METHOD: TWO OPT546 ADULT NASAL CANNULAS WERE RETURNED TO FPH IN (B)(4) FOR INSPECTION. BOTH WERE VISUALLY INSPECTED FOR DAMAGE. RESULTS: NO FAULT WAS FOUND WITH THE FIRST COMPLIANT DEVICE, WHICH WAS STILL SEALED WHEN RECEIVED. HOWEVER, VISUAL INSPECTION OF THE THE SECOND COMPLAINT DEVICE, WHICH WAS RETURNED UNSEALED, REVEALED THAT THE CANNULA MANIFOLD WAS NOT FULLY SEATED IN THE NASAL INTERFACE. A LOT CHECK REVEALED TWO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 100921. CONCLUSION: THE OPT546 NASAL CANNULA IS USED TO DELIVER HUMIDIFIED OXYGEN TO PATIENTS. IT CONSISTS OF A LIGHTWEIGHT DELIVERY TUBE WHICH IS CONNECTED TO A RIGID PLASTIC BASE AND SOFT NASAL PRONGS (NASAL INTERFACE). THE REPORTED FAULT CAN BE CAUSED BY THE SILICON BUCKLE BEING TIGHTENED TOO FIRMLY WHEN THE HEAD STRAP IS ADJUSTED, CAUSING THE CANNULA MANIFOLD TO BECOME LOOSE AND THE NASAL PRONGS TO COME APART FROM THE MANIFOLD. THE OPT546 NASAL CANNULA IS HELD IN PLACE BY A HEAD STRAP AND INCLUDES A LANYARD, WHICH IS PLACED AROUND THE PATIENT'S NECK OR ATTACHED TO THE PATIENT'S CLOTHING OR BEDDING TO REMOVE THE LOAD OF THE BREATHING CIRCUIT FROM THE PATIENT'S NARES. IF THE LANYARD IS NOT USED DURING THERAPY, THE CANNULA MANIFOLD WILL BE MORE SUSCEPTIBLE TO A PULLING FORCE AND EVENTUALLY BECOME LOOSE FROM THE NASAL INTERFACE. (B)(4).
A HEALTHCARE FACILITY IN (B)(6) REPORTED THE FOLLOWING TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE INVOLVING AN OPT546 ADULT NASAL CANNULA: "IF SOME TRACTIONS ARE APPLIED ON THE NASAL CANNULA'S, THEN THE SOFT PART SEPARATED FROM THE HARD PART OF THE NASAL PRONG. THE FLOW WHICH IS INTENDED IN THE NOSE, IS GOING NOT INTO THE NOSE BUT ESCAPES ALONG THE NOSE. THIS WAY AND WE WILL MISS THERAPY". NO PATIENT CONSEQUENCE WAS REPORTED AS A RESULT OF THIS INCIDENT.
A HEALTHCARE FACILITY IN (B)(6) REPORTED THE FOLLOWING TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE INVOLVING AN OPT546 ADULT NASAL CANNULA: "IF SOME TRACTIONS ARE APPLIED ON THE NASAL CANNULA'S, THEN THE SOFT PART SEPARATED FROM THE HARD PART OF THE NASAL PRONG. THE FLOW WHICH IS INTENDED IN THE NOSE, IS GOING NOT INTO THE NOSE BUT ESCAPES ALONG THE NOSE. THIS WAY AND WE WILL MISS THERAPY." NO PATIENT CONSEQUENCE WAS REPORTED AS A RESULT OF THIS INCIDENT.