FDA Adverse Event Death Summary report: N

MONITORING SERVICE APPLICATION (MSA)

MDR report key: 20541509 · Received October 27, 2024

Report

Report Number
3003584029-2024-00057
Event Type
Death
Date Received
October 27, 2024
Date of Event
May 22, 2024
Report Date
January 17, 2025
Manufacturer
BRAEMAR MANUFACTURING, LLC
Product Code
DQK
Removal / Correction Number
Z-0708-2025
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

FAILURE TO INFORM. THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 TRANSMITS ECG TRACING(S) TO A MONITORING APPLICATION FOR PRELIMINARY REVIEW BY A CARDIOLOGY TECHNICIAN AND SUBSEQUENT NOTIFICATION OR REPORTING TO THE TREATING CLINICIAN BASED ON THE TECHNICIAN FINDINGS. THE INVESTIGATION CONCLUDED THAT THE MONITORING APPLICATION DID NOT PRESENT AN ECG TRACING(S) CONTAINING AN ARRHYTHMIC EVENT THAT MET CLINICIAN-DEFINED NOTIFICATION CRITERIA TO A CARDIOLOGY TECHNICIAN, AND, THEREFORE, THE HEALTHCARE PROVIDER (HCP) WAS NOT NOTIFIED. NO ADDITIONAL INFORMATION REGARDING THIS EVENT IS KNOWN. NEITHER THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 NOR THE MONITORING APPLICATION PROVIDE A DIAGNOSIS OR TREATMENT RECOMMENDATION. INSTEAD, THE HCP DETERMINES A TREATMENT PLAN FOR THE PATIENT BASED ON THE INFORMATION PROVIDED. THIS SYSTEM IS NOT INTENDED FOR USE AS AN EMERGENCY RESPONSE SYSTEM FOR PATIENTS WHO MAY EXPERIENCE SERIOUS OR LIFE-THREATENING ARRHYTHMIAS. THE CAUSE OF THE PATIENT DEATH AND THE ROLE OF THE SYSTEM IN THE PATIENT'S DEATH, IF ANY, IS UNKNOWN. MISSED ARRHYTHMIC EVENT ON DAILY REPORT. THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 TRANSMITS ECG TRACING(S) TO A MONITORING APPLICATION FOR PRELIMINARY REVIEW BY A CARDIOLOGY TECHNICIAN AND SUBSEQUENT NOTIFICATION OR REPORTING TO THE TREATING CLINICIAN BASED ON THE TECHNICIAN FINDINGS. THE INVESTIGATION CONCLUDED THAT THE MONITORING APPLICATION DID NOT PRESENT AN ECG TRACING(S) CONTAINING AN ARRHYTHMIC EVENT THAT MET CRITERIA FOR REPRESENTATION ON THE DAILY AND END OF SERVICE REPORTS TO A CARDIOLOGY TECHNICIAN. THIS EVENT(S) DID NOT MEET CLINICIAN-DEFINED NOTIFICATION CRITERIA. THE ARRHYTHMIC EVENT(S) WAS NOT INCLUDED ON THE DAILY AND END OF SERVICE REPORTS PROVIDED TO THE HEALTHCARE PROVIDER (HCP). NO ADDITIONAL INFORMATION REGARDING THIS EVENT IS KNOWN. NEITHER THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 NOR THE MONITORING APPLICATION PROVIDE A DIAGNOSIS OR TREATMENT RECOMMENDATION. INSTEAD, THE HCP DETERMINES A TREATMENT PLAN FOR THE PATIENT BASED ON THE INFORMATION PROVIDED. THIS SYSTEM IS NOT INTENDED FOR USE AS AN EMERGENCY RESPONSE SYSTEM FOR PATIENTS WHO MAY EXPERIENCE SERIOUS OR LIFE-THREATENING ARRHYTHMIAS. THE CAUSE OF THE PATIENT DEATH IS UNKNOWN, BUT THE INVESTIGATION DID NOT IDENTIFY ANY EVIDENCE TO SUGGEST THE SYSTEM WAS A FACTOR. THIS REPORT IS BEING SUBMITTED AS REQUIRED BY FEDERAL REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, BRAEMAR MANUFACTURING, LLC, OR OUR EMPLOYEES, HAVE CAUSED OR CONTRIBUTED TO THE EVENT. FURTHERMORE, THIS REPORT REFLECTS THE EVENT AS ALLEGED WHEN REPORTED TO BRAEMAR MANUFACTURING, LLC AND DOES NOT IMPLY OR MEAN THAT THE INFORMATION AS REPORTED HEREIN IS ACCURATE OR HAS BEEN CONFIRMED BY BRAEMAR MANUFACTURING, LLC.

Additional Manufacturer Narrative · 0

FAILURE TO INFORM THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 TRANSMITS ECG TRACING(S) TO A MONITORING APPLICATION FOR PRELIMINARY REVIEW BY A CARDIOLOGY TECHNICIAN AND SUBSEQUENT NOTIFICATION OR REPORTING TO THE TREATING CLINICIAN BASED ON THE TECHNICIAN FINDINGS. THE INVESTIGATION CONCLUDED THAT THE MONITORING APPLICATION DID NOT PRESENT AN ECG TRACING(S) CONTAINING AN ARRHYTHMIC EVENT THAT MET CLINICIAN-DEFINED NOTIFICATION CRITERIA TO A CARDIOLOGY TECHNICIAN, AND, THEREFORE, THE HEALTHCARE PROVIDER (HCP) WAS NOT NOTIFIED. NO ADDITIONAL INFORMATION REGARDING THIS EVENT IS KNOWN. NEITHER THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 NOR THE MONITORING APPLICATION PROVIDE A DIAGNOSIS OR TREATMENT RECOMMENDATION. INSTEAD, THE HCP DETERMINES A TREATMENT PLAN FOR THE PATIENT BASED ON THE INFORMATION PROVIDED. THIS SYSTEM IS NOT INTENDED FOR USE AS AN EMERGENCY RESPONSE SYSTEM FOR PATIENTS WHO MAY EXPERIENCE SERIOUS OR LIFE-THREATENING ARRHYTHMIAS. THE CAUSE OF THE PATIENT DEATH AND THE ROLE OF THE SYSTEM IN THE PATIENT'S DEATH, IF ANY, IS UNKNOWN. MISSED ARRHYTHMIC EVENT ON DAILY REPORT THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 TRANSMITS ECG TRACING(S) TO A MONITORING APPLICATION FOR PRELIMINARY REVIEW BY A CARDIOLOGY TECHNICIAN AND SUBSEQUENT NOTIFICATION OR REPORTING TO THE TREATING CLINICIAN BASED ON THE TECHNICIAN FINDINGS. THE INVESTIGATION CONCLUDED THAT THE MONITORING APPLICATION DID NOT PRESENT AN ECG TRACING(S) CONTAINING AN ARRHYTHMIC EVENT THAT MET CRITERIA FOR REPRESENTATION ON THE DAILY AND END OF SERVICE REPORTS TO A CARDIOLOGY TECHNICIAN. THIS EVENT(S) DID NOT MEET CLINICIAN-DEFINED NOTIFICATION CRITERIA. THE ARRHYTHMIC EVENT(S) WAS NOT INCLUDED ON THE DAILY AND END OF SERVICE REPORTS PROVIDED TO THE HEALTHCARE PROVIDER (HCP). NO ADDITIONAL INFORMATION REGARDING THIS EVENT IS KNOWN. NEITHER THE BRAEMAR TELEMETRY PATCH SYSTEM BTPS-1000 NOR THE MONITORING APPLICATION PROVIDE A DIAGNOSIS OR TREATMENT RECOMMENDATION. INSTEAD, THE HCP DETERMINES A TREATMENT PLAN FOR THE PATIENT BASED ON THE INFORMATION PROVIDED. THIS SYSTEM IS NOT INTENDED FOR USE AS AN EMERGENCY RESPONSE SYSTEM FOR PATIENTS WHO MAY EXPERIENCE SERIOUS OR LIFE-THREATENING ARRHYTHMIAS. THE CAUSE OF THE PATIENT DEATH IS UNKNOWN, BUT THE INVESTIGATION DID NOT IDENTIFY ANY EVIDENCE TO SUGGEST THE SYSTEM WAS A FACTOR. THIS REPORT IS BEING SUBMITTED AS REQUIRED BY FEDERAL REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, BRAEMAR MANUFACTURING, LLC, OR OUR EMPLOYEES, HAVE CAUSED OR CONTRIBUTED TO THE EVENT. FURTHERMORE, THIS REPORT REFLECTS THE EVENT AS ALLEGED WHEN REPORTED TO BRAEMAR MANUFACTURING, LLC AND DOES NOT IMPLY OR MEAN THAT THE INFORMATION AS REPORTED HEREIN IS ACCURATE OR HAS BEEN CONFIRMED BY BRAEMAR MANUFACTURING, LLC. FOLLOW UP REPORT #1 SUBMITTED TO ENTER THE RECALL NUMBER ASSIGNED BY FDA.

Description of Event or Problem · 0

FAILURE TO INFORM. THE PATIENT EXPERIENCED AN ARRHYTHMIC EVENT(S) THAT MET CLINICIAN-DEFINED NOTIFICATION CRITERIA THAT WAS NOT COMMUNICATED DURING THE MONITORING PERIOD. BRAEMAR MANUFACTURING, LLC BECAME AWARE OF THE PATIENT DEATH IN ACCORDANCE WITH STANDARD PRACTICES AT THE END OF THE MONITORING PERIOD. MISSED ARRHYTHMIC EVENT ON DAILY REPORT. THE PATIENT EXPERIENCED AN ARRHYTHMIC EVENT(S) THAT WAS NOT DOCUMENTED AS A REPRESENTATIVE FINDING ON THE PATIENT'S DAILY AND END OF SERVICE REPORTS. THIS ARRHYTHMIC EVENT(S) DID NOT MEET CLINICIAN-DEFINED NOTIFICATION CRITERIA BUT SHOULD HAVE BEEN INCLUDED ON THE PATIENT'S DAILY AND END OF SERVICE REPORTS. BRAEMAR MANUFACTURING, LLC BECAME AWARE OF THE PATIENT DEATH IN ACCORDANCE WITH STANDARD PRACTICES AT THE END OF THE MONITORING PERIOD.

Description of Event or Problem · 0

FAILURE TO INFORM THE PATIENT EXPERIENCED AN ARRHYTHMIC EVENT(S) THAT MET CLINICIAN-DEFINED NOTIFICATION CRITERIA THAT WAS NOT COMMUNICATED DURING THE MONITORING PERIOD. BRAEMAR MANUFACTURING, LLC BECAME AWARE OF THE PATIENT DEATH IN ACCORDANCE WITH STANDARD PRACTICES AT THE END OF THE MONITORING PERIOD. MISSED ARRHYTHMIC EVENT ON DAILY REPORT THE PATIENT EXPERIENCED AN ARRHYTHMIC EVENT(S) THAT WAS NOT DOCUMENTED AS A REPRESENTATIVE FINDING ON THE PATIENT'S DAILY AND END OF SERVICE REPORTS. THIS ARRHYTHMIC EVENT(S) DID NOT MEET CLINICIAN-DEFINED NOTIFICATION CRITERIA BUT SHOULD HAVE BEEN INCLUDED ON THE PATIENT'S DAILY AND END OF SERVICE REPORTS. BRAEMAR MANUFACTURING, LLC BECAME AWARE OF THE PATIENT DEATH IN ACCORDANCE WITH STANDARD PRACTICES AT THE END OF THE MONITORING PERIOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1820279 MONITORING SERVICE APPLICATION (MSA) MONITORING SERVICE APPLICATION (MSA) DQK BRAEMAR MANUFACTURING, LLC

Patients

Seq Age Sex Outcome Treatment
1 83 YR Male Death