BD NEXIVA SP WITH MAXZERO
Report
- Report Number
- 1710034-2024-01212
- Event Type
- Malfunction
- Date Received
- October 27, 2024
- Date of Event
- August 9, 2024
- Report Date
- December 10, 2024
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- UDI-DI
- 00382903835577
- PMA / PMN Number
- K183399
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 383557 AND LOT NUMBER 4152107. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.
H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
NO ADDITIONAL INFORMATION.
IT WAS REPORTED THAT BD NEXIVA SP WITH MAXZERO DIFFICULTY DISENGAGING THE NEEDLE RESULTING IN PIERCING THE CATHETER AND LEAKAGE OF BLOOD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE NEEDLE IS SUPER DIFFICULT TO RETRACT AND WILL ACTUALLY GRIND WHILE FORCING A RETRACTION. IN SOME INSTANCES, THE FORCE THAT HAS HAD TO BE USED TO RETRACT THE NEEDLE HAS CAUSED A PUNCTURE IN THE TUBING WHICH THEN LEAKS AND WE HAVE TO RE-STICK THE PATIENT. INJURIES OR ADVERSE EVENT: NO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2026609 | BD NEXIVA SP WITH MAXZERO | PERIPHERAL IV CATHETERS | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 4152107 | 00382903835577 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |