FDA Adverse Event Malfunction Summary report: N

BD NEXIVA SP WITH MAXZERO

MDR report key: 20540992 · Received October 27, 2024

Report

Report Number
1710034-2024-01207
Event Type
Malfunction
Date Received
October 27, 2024
Date of Event
October 4, 2024
Report Date
November 11, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903835577
PMA / PMN Number
K183399
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. IT IS UNCLEAR IF THE DISTRIBUTOR IS REPORTING 3 OCT THIS AS THE EVENT DATE OR THEIR AWARE DATE.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 383557 AND LOT NUMBER 4152107. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD NEXIVA SP WITH MAXZERO NEEDLE DISENGAGEMENT DIFFICULTY RESULTING IN PIERCED CATHETER AND LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: INCIDENT DESCRIPTION: DIFFICULT TO RETRACT. REPORTED ISSUE: THE NEEDLE IS SUPER DIFFICULT TO RETRACT AND WILL ACTUALLY GRIND WHILE FORCING A RETRACTION. IN SOME INSTANCES, THE FORCE THAT HAS HAD TO BE USED TO RETRACT THE NEEDLE HAS CAUSED A PUNCTURE IN THE TUBING WHICH THEN LEAKS AND WE HAVE TO RE-STICK THE PATIENT. INJURIES OR ADVERSE EVENT: NO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1819268 BD NEXIVA SP WITH MAXZERO PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4152107 00382903835577

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown