TRUSTEEL
Report
- Report Number
- 3003442380-2024-29497
- Event Type
- Injury
- Date Received
- October 25, 2024
- Date of Event
- June 25, 2024
- Report Date
- August 22, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244018426
- PMA / PMN Number
- K041545
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. THE INFORMATION IN THIS COMPLAINT (B)(6) HAS BEEN EVALUATED. THE BATCH 6003939 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. INVESTIGATION PROCESS OF THE COMPLAINT WAS CARRIED OUT IN ACCORDANCE WITH WORK INSTRUCTION (WI) GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3.0 AND WI GUIDANCE FOR FUNCTIONAL TESTING FOR COMPLAINTS AREA VERSION 2.0 FOR THE CODE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE. COMPLAINT INVESTIGATIONS. PHYSICAL SAMPLES WERE FORMALLY REQUESTED; HOWEVER, THEY WERE NOT PROVIDED. IN ORDER TO TEST THE PRODUCT, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: PACKING. LOT 6003939 WAS MANUFACTURED ACCORDING TO THE WI VERSION 108 IN THE LINE 7, ON 29/OCT2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 05/AUG/2025 AGAINST MALFUNCTION CODE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE AND LOT 6003939 AND OTHER 2 COMPLAINTS HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT 6003939 AND MALFUNCTION CODE. CONCLUSION SUMMARY OF THE RELATED EVENT: THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES ON 25-JUNE-2024, IT WAS REPORTED THAT THE PATIENT ENCOUNTERED INFUSION SET CANNULA KINKED EVENT WITHIN 3 HOURS AFTER INSERTION. THE BLOOD GLUCOSE WAS REPORTED +900 MG/DL DUE TO WHICH PATIENT WAS HOSPITALIZED AND TREATED WITH IV AND FLUIDS OF SALINE AND INSULIN. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. COMPANY DO NOT SEE BENT/KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1721980 | TRUSTEEL | UNO CONTACT DETACH G29 60/6TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1002833 | 6003939 | 05705244018426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Female | Hospitalization| R |