FDA Adverse Event Malfunction Summary report: N

CARE ASSIST

MDR report key: 2053655 · Received March 17, 2011

Report

Report Number
9615739-2011-00068
Event Type
Malfunction
Date Received
March 17, 2011
Date of Event
February 28, 2011
Report Date
February 28, 2011
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TECHNICIAN FOUND THE BED IN REPAIR SHOP. FOUND THAT THE BRAKES WERE NOT WORKING - BRAKES WOULD SET BUT NOT HOLD. REPLACED CASTER BRAKES TO RESOLVE THIS ISSUE.

Description of Event or Problem · 1

COMPLAINT ALLEGED THAT THE BRAKES WERE NOT WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARE ASSIST AC POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER P1170

Patients

Seq Age Sex Outcome Treatment
1