FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 2053597 · Received April 13, 2011

Report

Report Number
2649622-2011-05942
Event Type
Injury
Date Received
April 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. MAX SVC LEAD IMPEDANCE RISES FROM AN APPROXIMATE BASELINE OF 70 OHMS THE WEEK OF (B)(6) 2010 TO A HIGH OF 13192 OHMS THE WEEK OF (B)(6) 2010. V PACE LEAD IMPEDANCE RISES ABRUPTLY FROM 944 OHMS THE WEEK OF (B)(6) 2010 TO 16382 OHMS THE REMAINING WEEKS OF THE RECORD. MAX SVC LEAD IMPEDANCE VARIES BETWEEN BASELINE AND HIGH VALUE (13192 OHMS) ERRATICALLY OVER THE WEEKS BETWEEN (B)(6) 2010 AND EXPLANT. MULTIPLE SHORT V-V SENSED EVENTS OF < 220 MS ARE OBSERVED ON THE (B)(6) 2010. (47 EPISODES NST (NON-SUSTAINED TACHYCARDIA), 4 EPISODES VF (VENTRICULAR FIBRILLATION)) HIGH V-SIC (SHORT INTERVAL COUNT) OBSERVED WITH 35360 COUNTS ON THE LIFETIME COUNTER, THE BULK OF THESE COUNTS OCCURRING SINCE THE (B)(6) 2010. HIGH SIC CAN INDICATE THE PRESENCE OF NOISE OR INTERMITTENCY IN THE SYSTEM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD FRACTURED, WAS OVERSENSING AND HAD HIGH IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944 ASKU

Patients

Seq Age Sex Outcome Treatment
1 75 YR Hospitalization| R 7232CX IMPLANTABLE PACEMAKER/CARDIO/DEFIB