MENTOR MEMORYGEL BREAST IMPLANT
Report
- Report Number
- 1645337-2024-12547
- Event Type
- Injury
- Date Received
- October 25, 2024
- Date of Event
- September 24, 2024
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317000389
- PMA / PMN Number
- P030053
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: BREAST PAIN, PTOSIS, MATERIAL RUPTURE, CAPSULAR CONTRACTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON NOVEMBER 15, 2024, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON NOVEMBER 20, 2024, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SM MPP GEL 800CC BREAST IMPLANT WAS RUPTURED. IN ADDITION, SHELL ABRASION WAS OBSERVED AT THE EDGES OF THE RUPTURE. THE EVALUATION DETERMINED THAT THE POSSIBLE CAUSE OF THE RUPTURE IS CONSISTENT WITH NORMAL WEAR. SHELL ABRASION SUGGESTS IN-VIVO FOLDING OR CREASING OF THE DEVICE. THIS MAY BE THE RESULT OF THE FOLLOWING FACTORS: CONTINUOUS AND SUSTAINED STRESSES TO THE DEVICE SUCH AS TOO SMALL BREAST POCKET AND FOLDING OR WRINKLING OF THE SHELL IN THE BREAST POCKET. IN SOME CASES, THE BREAST IMPLANTS MAY ALSO WEAR OUT OVER TIME. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST AUGMENTATION REVISION WITH TWO 800CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT SUFFERED A ZINGING SENSATION ON THE RIGHT BREAST, BILATERAL BREAST PAIN, AND BILATERAL BREAST PTOSIS. IN ADDITION, THE PATIENT WAS ALSO FOUND TO HAVE A LEFT BREAST CAPSULAR CONTRACTURE (BAKER GRADE IV) AND A RUPTURED LEFT BREAST IMPLANT. AS A RESULT, THE PATIENT UNDERWENT BILATERAL BREAST IMPLANT REMOVAL AND REPLACEMENT SURGERY ON (B)(6) 2024. THE REPLACEMENT DEVICES WERE: (LEFT) 790CC MENTOR MEMORYGEL XTRA BREAST IMPLANT CATALOG: SHPX790 LOT: 9993305 SN: (B)(6) AND (RIGHT) 790CC MENTOR MEMORYGEL XTRA BREAST IMPLANT CATALOG: SHPX790 LOT: 9993305 SN: (B)(6). THIS MEDWATCH FORM IS FOR THE LEFT BREAST PROSTHESIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1093204 | MENTOR MEMORYGEL BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 6422352 | 00081317000389 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Female | Required Intervention |