FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2053522 · Received April 13, 2011

Report

Report Number
2649622-2011-05923
Event Type
Injury
Date Received
April 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S LEAD INTEGRITY ALERT TRIGGERED. OVERSENSING WAS FOUND ON THE LEFT VENTRICULAR LEAD. AN APPARENT LEAD FRACTURE WAS SUSPECTED. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED. THE HIGH VOLTAGE COIL REMAINS IN USE. A NEW PACE/SENSE LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization| R 4194 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4076 IMPLANTABLE PACING LEAD