FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2053488 · Received April 13, 2011

Report

Report Number
2649622-2011-05914
Event Type
Injury
Date Received
April 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS T WAVE OVERSENSING ON THE RIGHT VENTRICULAR LEAD AS WELL AS OTHER SENSING DIFFICULTIES. THE POLARITY WAS CHANGED TO INTEGRATED BIPOLAR AND THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935 ASKU

Patients

Seq Age Sex Outcome Treatment
1 38 YR Required Intervention (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB