FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2053410 · Received April 13, 2011

Report

Report Number
2649622-2011-05897
Event Type
Injury
Date Received
April 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
Z0474-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT THE DEFIBRILLATOR CONDUCTOR WAS DISTORTED, THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION HAD COSMETIC DEPRESSION, AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD RISING AND VARYING IMPEDANCES. THE LEAD WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R