FDA Adverse Event Injury Summary report: N

SPRINT FIDELIS

MDR report key: 2053378 · Received April 13, 2011

Report

Report Number
2649622-2011-05886
Event Type
Injury
Date Received
April 13, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S029
Removal / Correction Number
Z-0067-0070-2008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. EVALUATION SUMMARY: (B)(4) THE DISTAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND THE DISTAL CONDUCTOR WAS FRACTURED. IT WAS ALSO NOTED THAT THE DEFIBRILLATOR CONDUCTOR WAS DISTORTED AND HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER TUBING OVERLAY WAS MELTED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE TIP ELECTRODE HAD A WHITE SUBSTANCE, AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS DISPLACED PROXIMALLY. INTERROGATION REVEALED THERE WAS POOR SENSING AND NO CAPTURE FROM THE LEAD. IT WAS ALSO NOTED THAT THE LEAD HAD HIGH THRESHOLDS AND AN APPARENT LEAD FRACTURE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT FIDELIS IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6949 ASKU

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization| O| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD