FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2053296 · Received April 13, 2011

Report

Report Number
2649622-2011-05845
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
February 8, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THERE WAS A TIP SEAL OBSERVATION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE HELIX OF THE LEAD WAS UNABLE TO DEPLOY. IN THE FIRST PLACEMENT ATTEMPT, THE HELIX TOOK FOURTEEN TURNS TO DEPLOY. ON THE SECOND PLACEMENT ATTEMPT, THE HELIX TOOK TWENTY TIMES AND WOULD NOT DEPLOY. THE LEAD WAS NOT USED AND A NEW LEAD IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE HELIX OF THE LEAD WAS UNABLE TO DEPLOY. IN THE FIRST PLACEMENT ATTEMPT, THE HELIX TOOK FOURTEEN TURNS TO DEPLOY. ON THE SECOND PLACEMENT ATTEMPT, THE HELIX TOOK TWENTY TIMES AND WOULD NOT DEPLOY. THE LEAD WAS NOT USED AND A NEW LEAD IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other (B)(4) IMPLANTABLE PACING LEAD