FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2053182 · Received April 13, 2011

Report

Report Number
2649622-2011-05805
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
February 1, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY FOR (B)(4): NO ANOMALIES FOUND. FULL LEAD RETURNED AND ANALYZED. ADDITIONAL FINDING OF INNER TUBING KINKED/BUCKLED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR LEAD SHOWED NOISE. THE CABLES, ADAPTOR, AND PROGRAMMER WERE CHANGED BUT THE NOISE PERSISTED. THE LEAD WAS NOT IMPLANTED AND REPLACED BY ANOTHER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 63 YR Other