FDA Adverse Event Injury Summary report: N

PROTECTA XT DR

MDR report key: 2053168 · Received April 13, 2011

Report

Report Number
6000094-2011-00528
Event Type
Injury
Date Received
April 13, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
NIK
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED OR CONFIRMED MALFUNCTION.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT RECEIVED INAPPROPRIATE SHOCKS DUE TO LEAD NOISE CAUSED BY A LOOSE SET SCREW ON THE DEVICE. THE SET SCREW WAS TIGHTENED AND THE DEVICE REMAINS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.