FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2053164 · Received April 13, 2011

Report

Report Number
2649622-2011-05795
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
February 1, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE LEAD IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY FOR (B)(4): HELIX DISENGAGED FROM HELICAL CHANNEL. FULL LEAD RETURNED AND ANALYZED. ADDITIONAL FINDINGS OF DISTAL CONDUCTOR BLOOD/BODY FLUID (NOT OBSTRUCTED), BLOOD IN/ON HELIX MECHANISM (SLEEVE HEAD) AND IN/ON HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HELIX GOT STUCK IN THE RETRACTED POSITION WHEN THE PHYSICIAN WAS ATTEMPTING TO REPOSITION THE LEAD. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other