CARDIOSAVE HYBRID, TYPE B PLUG
Report
- Report Number
- 2249723-2024-0004373
- Event Type
- Malfunction
- Date Received
- October 25, 2024
- Date of Event
- October 4, 2024
- Report Date
- November 14, 2024
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- UDI-DI
- 10607567108391
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
UPDATED FIELDS - B4, D9, G3, G6, H2, H3, H4, H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, COMPONENT CODES, INVESTIGATION CONCLUSIONS), H11. A GETINGE FIELD SERVICE ENGINEER (FSE) EVALUATED THE UNIT AND ALSO FOUND FAULT LOGS 77 IN THE RECENT FAULTS. THE FSE REPLACED SCROLL COMPRESSOR (0119-00-0236), MUFFLER <100 (0103-00-0499) AND MUFFLER 1/8 (0103-00-0631). AFTER FURTHER TESTING THE IABP THE UNIT ALSO FAILED THE PIM TEST. REPLACED PNEUMATIC MODULE (0997-00-1178) AND PIM TEST PASSED SUCCESSFULLY. PERFORMED A PM, FULL CALIBRATION, AND TESTED THE UNIT. THE PRODUCT PASSED ALL FUNCTIONAL/SAFETY TESTING. RETURNED THE UNIT TO THE CUSTOMER.
IT WAS REPORTED THAT PRIOR TO USE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD A POWER UP TEST FAILURE CODE 14. THERE WAS NO PATIENT INVOLVEMENT.
N/A
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1092955 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. | 0998-00-0800-53 | 10607567108391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | N/A. |