FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID, TYPE B PLUG

MDR report key: 20531402 · Received October 25, 2024

Report

Report Number
2249723-2024-0004373
Event Type
Malfunction
Date Received
October 25, 2024
Date of Event
October 4, 2024
Report Date
November 14, 2024
Manufacturer
DATASCOPE CORP.
Product Code
DSP
UDI-DI
10607567108391
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

UPDATED FIELDS - B4, D9, G3, G6, H2, H3, H4, H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, COMPONENT CODES, INVESTIGATION CONCLUSIONS), H11. A GETINGE FIELD SERVICE ENGINEER (FSE) EVALUATED THE UNIT AND ALSO FOUND FAULT LOGS 77 IN THE RECENT FAULTS. THE FSE REPLACED SCROLL COMPRESSOR (0119-00-0236), MUFFLER <100 (0103-00-0499) AND MUFFLER 1/8 (0103-00-0631). AFTER FURTHER TESTING THE IABP THE UNIT ALSO FAILED THE PIM TEST. REPLACED PNEUMATIC MODULE (0997-00-1178) AND PIM TEST PASSED SUCCESSFULLY. PERFORMED A PM, FULL CALIBRATION, AND TESTED THE UNIT. THE PRODUCT PASSED ALL FUNCTIONAL/SAFETY TESTING. RETURNED THE UNIT TO THE CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD A POWER UP TEST FAILURE CODE 14. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

N/A

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1092955 CARDIOSAVE HYBRID, TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. 0998-00-0800-53 10607567108391

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown N/A.