FDA Adverse Event
Injury
Summary report: N
FINELINE II
MDR report key: 2053105
·
Received April 13, 2011
Report
- Report Number
- 2124215-2011-04042
- Event Type
- Injury
- Date Received
- April 13, 2011
- Date of Event
- March 2, 2011
- Report Date
- March 2, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE AND MINIMAL SENSING. UPON OPENING THE POCKET DURING THE REVISION PROCEDURE, THE RV LEAD WAS FOUND TO BE AT 90 DEGREE ANGLE DUE TO A FRACTURE. THE LEAD WAS SURGICALLY ABANDONED AND A NEW LEAD WAS SUCCESSFULLY IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4456 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| L| R | 4469| 1290| 4456| 1190 |