FDA Adverse Event Injury Summary report: N

MARQUIS DR

MDR report key: 2053097 · Received April 13, 2011

Report

Report Number
6000094-2011-00523
Event Type
Injury
Date Received
April 13, 2011
Date of Event
January 1, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
P980016/S23
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE SET SCREW WAS LOOSE/DETACHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WERE SENSING DIFFICULTIES DUE TO A DEFECTIVE SETSCREW ON THE DEVICE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS REPORTED THAT THERE WERE DIFFICULTIES ATTACHING A LEAD DUE TO A PROBLEM WITH THE SETSCREW. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MARQUIS DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. 7274 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R