FDA Adverse Event Injury Summary report: N

COGNIS

MDR report key: 2053049 · Received April 13, 2011

Report

Report Number
2124215-2011-04124
Event Type
Injury
Date Received
April 13, 2011
Date of Event
March 2, 2011
Report Date
May 7, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE REMAINS IMPLANTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THE DEVICE REMAINS IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT DUE TO GREATER THAN 200 OHM SHOCK LEAD IMPEDANCE MEASUREMENTS, A LEAD EXTRACTION WAS PERFORMED. DURING THE REVISION PROCEDURE, A VENOGRAM SHOWED VENOUS OCCLUSION. IT WAS NOTED UPON INSPECTION THAT THE TERMINAL END FOR THE DISTAL COIL WAS NOT FASTENED INTO THE HEADER, THE SET SCREW WAS LOOSE AND THE LEAD PULLED OUT EASILY. THE RV LEAD WAS RE-SEATED INTO THE HEADER WITH SUCCESSFUL DEFIBRILLATION THRESHOLD TESTING WITH NORMAL SHOCK LEAD IMPEDANCE MEASUREMENTS (31 OHMS). THE LEAD REMAINED IN THE TRIAD CONFIGURATION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT INCREASED SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS WERE OBSERVED WITH THIS DEVICE SYSTEM. TECHNICAL SERVICES DISCUSSED EVALUATION RECOMMENDATIONS. SEVERAL PAINLESS SHOCK IMPEDANCE TESTS WERE PERFORMED WITH GREATER THAN 125 OHMS MEASURED. NON-INVASIVE PROGRAMMED STIMULATION (NIPS) TESTING WAS PERFORMED AND THE DEVICE CONVERTED SUCCESSFULLY AT 21 J. THE PAINLESS TEST WAS AGAIN PERFORMED, RESULTING IN IMPEDANCE MEASUREMENTS WITHIN ACCEPTABLE LIMITS. THE PHYSICIAN ELECTED TO CONTINUE TO MONITOR THE DEVICE SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| L| R N119| H177| 5076| 4543| 1853| 0158