FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 2053037 · Received April 13, 2011

Report

Report Number
2649622-2011-05747
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
January 25, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE LEAD WAS TOO SMALL FOR THE VESSEL. THE INSULATION APPEARED TO BE DENTED AND WAS REPAIRED. THE LEAD HAD UNACCEPTABLE THRESHOLDS AND THE POSITION WAS QUESTIONABLE. THE LEAD WAS NOT IMPLANTED AND REPLACED WITH A NEW ONE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other