FDA Adverse Event Malfunction Summary report: N

BD BBL¿ TRYPTICASE¿ SOY BROTH

MDR report key: 20530051 · Received October 24, 2024

Report

Report Number
1119779-2024-00783
Event Type
Malfunction
Date Received
October 24, 2024
Date of Event
October 4, 2024
Report Date
March 10, 2025
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JSG
UDI-DI
30382902210932
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: MATERIAL 221093 IS MANUFACTURED BY REHYDRATING THE MEDIA COMPONENTS WITH USP PURIFIED WATER AND THOROUGHLY MIXING UNTIL A HOMOGENEOUS SOLUTION IS OBTAINED. THE TUBES ARE FILLED, CAPPED, TORQUED, AND THEN LABELED BY MACHINE PER STANDARD OPERATING PROCEDURE (SOP). THE TUBES ARE TERMINALLY AUTOCLAVED IN AN AIR OVER PRESSURE (AOP) AUTOCLAVE, PER MANUFACTURING INSTRUCTIONS, USING A VALIDATED CYCLE. POST AUTOCLAVING, TUBES ARE PACKAGED INTO FINAL SHIPPING CONFIGURATIONS. THE BATCH HISTORY RECORD REVIEW FOR BATCH 4123398 WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. FORMULATION, FILLING, TORQUING, AND AUTOCLAVING PROCESSES WERE WITHIN SPECIFICATIONS. QC INSPECTION AND TESTING WERE SATISFACTORY AT TIME OF RELEASE. AS PART OF THE RELEASE CRITERIA FOR THIS PRODUCT, THE BHR IS REVIEWED TO CONFIRM THE FOLLOWING: THE TOTAL ELAPSED TIME BETWEEN END OF FORMULATION AND START OF THE AUTOCLAVE CYCLE WAS WITHIN THE SPECIFIED LIMITS. ALL AUTOCLAVE PARAMETERS CONFORMED TO THE VALIDATED CYCLE PARAMETERS FOR THIS PRODUCT. THE MINIMUM F0 FOR THIS PRODUCT WAS MET. THE COMPLAINT HISTORY WAS REVIEWED, AND OTHER COMPLAINTS HAVE BEEN TAKEN ON THIS BATCH FOR CONTAMINATION. RETENTION SAMPLES FROM BATCH 4123398 WERE NOT AVAILABLE FOR INSPECTION. NO RETURN SAMPLES OR PHOTOS WERE RECEIVED FOR INVESTIGATION OF THIS COMPLAINT. THIS COMPLAINT CANNOT BE CONFIRMED FOR CONTAMINATION.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH THAT GRAM-NEGATIVE BACILLI WERE OBSERVED IN A GRAM STAIN OF A CSF CULTURE AFTER 5 DAYS INCUBATION AT 35°C. THE END USER SUSPECTED CONTAMINATION AND PERFORMED A GRAM STAIN ON UNUSED SOY BROTH. GRAM-NEGATIVE BACILLI WERE OBSERVED IN AN UNSPECIFIED NUMBER OF TUBES. SUBCULTURE OF THE BROTH RESULTED IN NO GROWTH. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD BBL¿ TRYPTICASE¿ SOY BROTH THAT GRAM-NEGATIVE BACILLI WERE OBSERVED IN A GRAM STAIN OF A CSF CULTURE AFTER 5 DAYS INCUBATION AT 35°C. THE END USER SUSPECTED CONTAMINATION AND PERFORMED A GRAM STAIN ON UNUSED SOY BROTH. GRAM-NEGATIVE BACILLI WERE OBSERVED IN AN UNSPECIFIED NUMBER OF TUBES. SUBCULTURE OF THE BROTH RESULTED IN NO GROWTH. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1818584 BD BBL¿ TRYPTICASE¿ SOY BROTH CULTURE MEDIA, NON-SELECTIVE AND NON-DIFFERENTIAL JSG BECTON DICKINSON & CO. (SPARKS) 4123398 30382902210932

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown