FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2052988 · Received April 13, 2011

Report

Report Number
2649622-2011-05730
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
February 7, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE LEAD WAS ATTEMPTED BUT UNABLE TO BE USED DUE TO THE PATIENT'S ANATOMY. ANOTHER LEAD WAS IMPLANTED INSTEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other