FDA Adverse Event
Injury
Summary report: N
EASYTRAK 2
MDR report key: 2052919
·
Received April 13, 2011
Report
- Report Number
- 2124215-2011-03930
- Event Type
- Injury
- Date Received
- April 13, 2011
- Date of Event
- February 28, 2011
- Report Date
- March 24, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, BOTH THE MECHANICAL AND THE ELECTRICAL PERFORMANCE OF THE LEAD UNDER COMPLAINT WERE SCRUTINIZED. THE ANALYSIS DID NOT SHOW ANY DEVIATIONS FROM THE ELECTRICAL SPECIFICATIONS. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
Additional Manufacturer Narrative · 1
AS OF THIS DATE, THE LEAD HAS NOT BEEN RETURNED. IF THIS LEAD SHOULD BE RETURNED TO OUR COMPANY, IT WILL BE ANALYZED AND THIS REPORT WILL BE REOPENED AND UPDATED AS NECESSARY.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT POST IMPLANT THIS LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4543 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention | 0185| 4136| H227| 4543 |