FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 2052919 · Received April 13, 2011

Report

Report Number
2124215-2011-03930
Event Type
Injury
Date Received
April 13, 2011
Date of Event
February 28, 2011
Report Date
March 24, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, BOTH THE MECHANICAL AND THE ELECTRICAL PERFORMANCE OF THE LEAD UNDER COMPLAINT WERE SCRUTINIZED. THE ANALYSIS DID NOT SHOW ANY DEVIATIONS FROM THE ELECTRICAL SPECIFICATIONS. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE LEAD HAS NOT BEEN RETURNED. IF THIS LEAD SHOULD BE RETURNED TO OUR COMPANY, IT WILL BE ANALYZED AND THIS REPORT WILL BE REOPENED AND UPDATED AS NECESSARY.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT POST IMPLANT THIS LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4543

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention 0185| 4136| H227| 4543