FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2052891 · Received April 13, 2011

Report

Report Number
2124215-2011-03388
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
February 24, 2011
Report Date
February 27, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED AND IS CURRENTLY BEING ANALYZED. THIS EVENT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST-MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THAT THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR UNKNOWN REASONS. NO DEVICE ALLEGATIONS WERE REPORTED. THIS EVENT WAS CREATED DUE TO THE INITIAL TESTING PERFORMED BY OUR POST MARKET QUALITY ASSURANCE LABORATORY. INITIAL TESTING NOTED A POSSIBLE PERFORMANCE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H175

Patients

Seq Age Sex Outcome Treatment
1 80 YR H175| 4244| 0185| 4517