ARTICULEZE M HEAD 36MM +1.5
Report
- Report Number
- 1818910-2011-05734
- Event Type
- Injury
- Date Received
- April 7, 2011
- Date of Event
- March 8, 2011
- Report Date
- March 8, 2011
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- JDI
- PMA / PMN Number
- K980513
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
FOLLOW-UP FOR ADDITIONAL EVENT INFORMATION WAS CONDUCTED UTILIZING WORK INSTRUCTION WI-7915 APPENDIX A; REV. A. THE NEWLY PROVIDED INFORMATION WAS REVIEWED. REVIEW OF THE SUPPLIED INFORMATION FOUND EVIDENCE OF HIGH ION LEVELS. THE ELEVATED ION LEVELS ARE LIKELY DUE TO IMPINGEMENT/ POOR DEVICE POSITIONING AND THE PATIENT HAVING BI-LATERAL HIP IMPLANTS WHICH COULD INCREASE THE LEVELS. NO EVIDENCE WAS FOUND SUGGESTING PRODUCT ERROR WAS A CONTRIBUTING FACTOR AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
PT WAS REVISED TO ADDRESS HIGH LEVELS OF METAL IONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARTICULEZE M HEAD 36MM +1.5 | 87JDI; 87LPH | JDI | DEPUY ORTHOPAEDICS, INC. | NA | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |