INSIGNIA
Report
- Report Number
- 2124215-2011-03673
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- February 25, 2011
- Report Date
- March 28, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION, BUT DECLARED END OF LIFE (EOL) EARLIER THAN PREVIOUSLY ESTIMATED. DECLARATION OF THIS INDICATOR OCCURS AFTER THE DEVICE COMPLETES AN INTERNAL CALCULATION OF BATTERY CONSUMPTION. FACTORS INFLUENCING THIS CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH, LEAD IMPEDANCE, AND THE LAST 30 DAYS AVERAGE PACING PERCENTAGE. ANY CHANGES IN THOSE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION. THE ESTIMATED LONGEVITY REMAINING AND BATTERY STATUS GAUGE DISPLAYED BY THE PROGRAMMER ARE BASED ON BATTERY CURRENT CONSUMPTION CALCULATIONS AT THE TIME OF INTERROGATION, AND CALCULATIONS PERFORMED INTERNAL TO THE DEVICE ARE SUSPENDED DURING THE SESSION. WHEN AMBULATORY (AWAY FROM THE PROGRAMMER), THE DEVICE PERIODICALLY ASSESSES OPERATING CURRENT AND MAY REVISE THE ELECTIVE REPLACEMENT TIME (ERT) DECLARATION POINT TO ENSURE A MINIMUM OF 3 MONTHS BATTERY LIFE BETWEEN ERT AND EOL. THIS DEVICE ASSESSMENT OF OPERATING CURRENT, BATTERY CHARGE STATE, AND REVISION OF THE ERT DECLARATION POINT MAY CAUSE DIFFERENCES AND FLUCTUATIONS RELATIVE TO PREVIOUS PROGRAMMER BATTERY STATUS INDICATORS.
AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED. IF THIS DEVICE SHOULD BE RETURNED TO OUR COMPANY, IT WILL BE ANALYZED AND THIS INVESTIGATION WILL BE REOPENED AND UPDATED AS NECESSARY.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE BATTERY ONE YEAR AGO STATED BEGINNING OF LIFE (BOL) AND NOW IS SHOWING END OF LIFE (EOL) INDICATING THAT THE BATTERY MAY BE DEPLETING PREMATURELY. THIS DEVICE IS PART OF MICROCRYSTAL FAILURE MODE 2 ADVISORY, HOWEVER NO ADVISORY INDICATIONS WERE PRESENT. THE DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | S606| 4087| 1290| MISMATCH| 1388TC| 1388T| 4469 |