UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-01090
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- March 14, 2011
- Report Date
- March 14, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER HAD PREVIOUSLY REPLACED THE CO2 MEMBRANE. THE CUSTOMER NOTED THAT IN THE DAYS PRIOR TO THE EVENT, THEY HAD TO RECALIBRATE CO2 MORE OFTEN. WHEN THE TREND WAS NOTICED, QC WAS RERUN AND WAS LOW. SERVICE REPLACED BOTH CO2 ELECTRODES AND VERIFIED PERFORMANCE.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO ERRONEOUSLY LOW CARBON DIOXIDE (CO2) RESULTS GENERATED BY UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. WHEN THE TREND WAS NOTICED, ELEVEN (11) SAMPLES WERE RERUN ON THE CUSTOMER'S OTHER INSTRUMENT AND 2 REPORTS WERE AMENDED. OF THE ELEVEN (11) SAMPLES, THE DIFFERENCE IN RESULTS FOR SEVEN (7) OF THE SAMPLES EXCEEDED ASSAY PRECISION CLAIMS. IT IS UNKNOWN WHICH TWO (2) SAMPLES WERE AMENDED. THE SEVEN (7) SAMPLES THAT EXCEEDED CLAIMS ARE SHOWN. PATIENT TREATMENT WAS NOT INITIATED OR WITHHELD BASED UPON THE FALSE RESULTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | CLINICAL CHEMSTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |