BD PHOENIX PANEL NMIC-306
Report
- Report Number
- 1119779-2024-00776
- Event Type
- Malfunction
- Date Received
- October 24, 2024
- Date of Event
- October 4, 2024
- Report Date
- December 3, 2024
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 30382904492923
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
D.4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H.4. DEVICE MANUFACTURE DATE: UNKNOWN. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, K031530, K031699, K031912, K032299, K032567, K032655, K033362, K033560, K041384, K042932, K052269, K060214, K060217, K060257, K060444, K060447, K061327, K061355, K062207, K062944, K063301, K063486, K063573, K063811, K063824, K071623, K132674, K132909, K151320, K181665, K173252, K190905, K163637, K173523, K033458, AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY: THIS COMPLAINT IS FOR FALSE POSITIVE CPO RESULTS WITH PSEUDOMONAS AERUGINOSA WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER UNKNOWN. THE CUSTOMER DID NOT PROVIDE PANEL OR ISOLATE RETURNS BUT PROVIDED PHOENIX GENERATED LAB REPORTS FOR THE INVESTIGATION. THE LAB REPORTS SHOW RESULTS FOR P. AERUGINOSA. THE BATCH NUMBER WAS NOT PROVIDED; THEREFORE, THIS COMPLAINT IS UNCONFIRMED. A REVIEW OF QUALITY NOTIFICATIONS COULD NOT BE PERFORMED SINCE A BATCH NUMBER WAS NOT PROVIDED. COMPLAINT TRENDING WAS PERFORMED, AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.
IT WAS REPORTED WHILE USING THE BD PHOENIX NMIC-306 A PATIENT ISOLATE HAD RESULTED AS CARBAPENEMASE-PRODUCING ORGANISM (CPO) POSITIVE WHILE THE CARBAPENEM DRUGS HAD RESULTED AS SENSITIVE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
IT WAS REPORTED WHILE USING THE BD PHOENIX NMIC-306 A PATIENT ISOLATE HAD RESULTED AS CARBAPENEMASE-PRODUCING ORGANISM (CPO) POSITIVE WHILE THE CARBAPENEM DRUGS HAD RESULTED AS SENSITIVE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1737731 | BD PHOENIX PANEL NMIC-306 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY | LON | BECTON DICKINSON & CO. (SPARKS) | UNKNOWN | 30382904492923 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |