FDA Adverse Event Malfunction Summary report: N

GLUCOMMANDER

MDR report key: 20526568 · Received October 24, 2024

Report

Report Number
3005853093-2024-00001
Event Type
Malfunction
Date Received
October 24, 2024
Report Date
October 15, 2024
Manufacturer
GLYTEC, LLC
Product Code
NDC
UDI-DI
00860057000305
PMA / PMN Number
K152300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

AN UNCOMMON WORKFLOW IN GLUCOMMANDER SUBQ V3.5.1.2 OR LATER, E.G., DELIVERING BASAL DOSE MANY HOURS LATE AND CHANGING THE DOSE DISTRIBUTION, MAY RESULT IN A PLANNED GUARDRAIL TO NOT FUNCTION AS INTENDED RESULTING IN A HIGHER THAN EXPECTED BASAL DOSE RECOMMENDATION. THIS WAS OBSERVED TO OCCUR ONCE IN 100S OF 1000S OF BASAL DOSE RECOMMENDATIONS. THIS ISSUE WAS ADDRESSED IN SOFTWARE PATCH 3.5.2.2, RELEASED APRIL 27, 2023 AND IMPLEMENTED BY THE CUSTOMER MAY 6, 2023. THIS MDR WAS INITIALLY SUBMITTED THROUGH ESG ON 08-FEB-2023 AS "FDA MDR 08-FEB-2023 FINAL VERSION.PDF". ON 25-SEP-2024, IT CAME TO GLYTEC'S ATTENTION THAT THE EMDR HAD FAILED TO LOAD SUCCESSFULLY INTO THE CDRH ADVERSE EVENT DATABASE. HENCE, GLYTEC CONTACTED CESUB HELP DESK ON 25-SEP-2024 TO INQUIRE WHY THE SUBMISSION ON 08-FEB-2023 FAILED. CESUB HELP LOGGED GLYTEC'S INQUIRY AS CSH-36933. ON 01-OCT-2024, CESUB RESPONDED TO GLYTEC STATING: "THE SUBMITTED PDF FILE AS THE SUBMISSION FILE BUT THE SUBMISSION FILE HAS TO BE A ZIP FILE THAT'S CREATED USING ESUBMITTER. PLEASE ADVISE THE USER TO USE ESUBMITTER TO CREATE THEIR SUBMISSION." THEREFORE, GLYTEC IS GOING TO RE-SUBMIT THIS MDR VIA ESG AFTER RE-CREATING IT (THIS REPORT YOU ARE READING NOW) USING ESUBMITTER.

Description of Event or Problem · 0

GLYTEC RECEIVED A CALL FROM A HEALTHCARE PROFESSIONAL WHO WAS ATTEMPTING TO MODIFY A PATIENT'S RECOMMENDED BASAL DOSAGES AND DOSE FREQUENCY. UPON DOING SO, THE NEW BASAL DOSE RECOMMENDATION WAS HIGHER THAN THE NURSE EXPECTED. THIS DOSE WAS NOT ADMINISTERED TO THE PATIENT AND NO ADVERSE EVENTS NECESSITATING MEDICAL INTERVENTION WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1856731 GLUCOMMANDER ELECTRONIC GLYCEMIC MANAGEMENT SYSTEM (EGMS) NDC GLYTEC, LLC 3.5.1.2 00860057000305

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown