FDA Adverse Event Injury Summary report: N

EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM

MDR report key: 20526450 · Received October 24, 2024

Report

Report Number
2015691-2024-08074
Event Type
Injury
Date Received
October 24, 2024
Date of Event
October 11, 2024
Report Date
December 11, 2024
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPW
UDI-DI
00690103212846
PMA / PMN Number
P230013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE EVENT IS CAPTURED BY EDWARDS LIFESCIENCES UNDER COMPLAINT #: (B)(4). THE MANUFACTURER'S INVESTIGATION IS ONGOING AND A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 0

THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED/ADDED: B4, G3, G6, H2, H6 AND H11. THE COMPLAINT FOR VALVE INTERACTS WITH CONDUCTION SYSTEM WAS CONFIRMED WITH OTHER EMPIRICAL EVIDENCE VIA INFORMATION PROVIDED BY EDWARDS CLINICAL SPECIALIST. THE REPORTED CONDUCTION DISTURBANCE DOES NOT ALLEGE A MALFUNCTION THAT COULD BE RELATED TO AN EDWARDS MANUFACTURING DEFICIENCY. THE EVENT DOES NOT ALLEGE A LABELING ISSUE OR DEVICE RELATED INFECTION; THEREFORE, NO DHR, LOT HISTORY REVIEW, OR MANUFACTURING MITIGATIONS REVIEW ARE REQUIRED. SINCE NO MANUFACTURING NON-CONFORMANCES WERE FOUND AND NO LABELING, TRAINING, OR IFU DEFICIENCIES WERE IDENTIFIED, NO PREVENTATIVE OR CORRECTIVE ACTIONS WERE REQUIRED.

Description of Event or Problem · 0

AS REPORTED BY THE EDWARDS LIFESCIENCES AFFILIATE IN THE UNITED KINGDOM, EDWARDS RECEIVED NOTIFICATION OF A 52MM EVOQUE PROCEDURE IN TRICUSPID POSITION BY RIGHT FEMORAL VEIN WHERE THE VALVE WAS IMPLANTED IN THE DESIRED POSITION. ON POST-OPERATIVE DAY 15 IT WAS LEARNT THAT A PERMANENT PACEMAKER WAS IMPLANTED. DURING THE PACEMAKER IMPLANTATION A PERFORATION WAS OBSERVED WITH THORACOTOMY. THE PATIENT WAS IN INTENSIVE CARE UNIT AND EXPERIENCED HEART FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2566982 EVOQUE TRICUSPID VALVE REPLACEMENT SYSTEM PERCUTANEOUSLY DELIVERED PROSTHESES AND TRICUSPID VALVES NPW EDWARDS LIFESCIENCES 9850EV52 00690103212846

Patients

Seq Age Sex Outcome Treatment
1 77 YR Unknown Required Intervention