FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 2052631 · Received April 13, 2011

Report

Report Number
2124215-2011-03439
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
February 24, 2011
Report Date
March 1, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

BOSTON SCIENTIFIC'S POST MARKET QUALITY ASSURANCE LABORATORY DETERMINED THAT THE DEVICE HAD NORMAL TELEMETRY OPERATIONS, AS WELL AS NORMAL PACING AND SENSING FUNCTIONS. A REVIEW OF THE DEVICE'S MEMORY REVEALED THAT A RATE FAULT RESET WAS STORED IN THE RESET COUNTER. LABORATORY TESTING HAS SHOWN THAT THE AUTO LEAD DETECT FEATURE CAN CAUSE A RATE FAULT RESET TO OCCUR, DUE TO THE MAX TRACKING RATE PROTECTION NOT BEING INITIALIZED. THIS TYPE OF RESET OCCURS PRIOR TO LEAD ATTACHMENT. ONCE A LEAD IS DETECTED, THE AUTO LEAD DETECT FEATURE WILL BE DISABLED, ALLOWING THE MAX TRACKING RATE PROTECTION TO BE INITIALIZED, PREVENTING THE RATE FAULT RESETS FROM OCCURRING. THIS RESET IS NORMAL DIAGNOSTIC BEHAVIOR AND DOES NOT HAVE ANY IMPACT ON THE THERAPY PROVIDED TO THE PATIENT. THE DEVICE PERFORMED NORMALLY THROUGHOUT ANALYSIS AND WAS FOUND TO BE FUNCTIONING AS DESIGNED.

Additional Manufacturer Narrative · 1

AT THIS TIME, THIS DEVICE IS UNDERGOING DETAILED ANALYSIS. WHEN ANALYSIS IS COMPLETE, THIS EVENT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED BECAUSE THE PATIENT NEEDED THEIR THERAPY UPGRADED DUE TO THE PATIENT'S CONDITION. THERE WAS NO ALLEGATION FROM THE FIELD REGARDING THE FUNCTION OF THE DEVICE. THIS EVENT WAS CREATED DUE TO THE INITIAL TESTING PERFORMED BY OUR POST MARKET QUALITY ASSURANCE LABORATORY. INITIAL TESTING NOTED A POSSIBLE PERFORMANCE ISSUE CONCERNING A RATE FAULT RESET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1194

Patients

Seq Age Sex Outcome Treatment
1 73 YR 4469| S603| 1194| 4470