FDA Adverse Event Injury Summary report: N

COMMUNICATOR

MDR report key: 2052587 · Received April 13, 2011

Report

Report Number
2124215-2011-03458
Event Type
Injury
Date Received
April 13, 2011
Date of Event
February 24, 2011
Report Date
August 8, 2011
Manufacturer
EXTERNAL MANUFACTURER
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PENDING PRODUCT RETURN AND ANALYSIS. THIS SECTION WILL BE UPDATED UPON COMPLETION OF THE LAB ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE POWER BRICK WAS DEFORMED FROM HEAT AND A BURNING SMELL WAS NOTED. THE POWER BRICK FAILED TO OUTPUT ANY SIGNIFICANT VOLTAGE WHILE PLUGGED INTO A POWER SOURCE. THE POWER BRICK'S CASE TEMPERATURE REACHED 109.1 DEGREES C. AFTER 25 MINUTES.

Additional Manufacturer Narrative · 1

THE POWER SUPPLY IS BEING ANALYZED BY BSC CRM POST MARKET QUALITY ASSURANCE LABORATORY. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THE POWER SUPPLY FOR THIS COMMUNICATOR WAS HOT TO THE TOUCH. THE PATIENT REPORTED THAT HE DID NOT RECEIVE A THIRD DEGREE BURN BUT THE POWER SUPPLY WAS VERY HOT THE TOUCH. NO MEDICAL OR SURGICAL INTERVENTION WAS REPORTED. THE COMMUNICATOR AND POWER SUPPLY WERE REPLACED. THE POWER SUPPLY WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC CRM RECEIVED INFORMATION THAT THE POWER CORD TO THIS COMMUNICATOR WAS HOT TO THE TOUCH AND BURNED THE PATIENT. AS A RESULT, THE COMMUNICATOR AND POWER CORD WERE REPLACED.