FDA Adverse Event Malfunction Summary report: N

COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH

MDR report key: 2052514 · Received April 13, 2011

Report

Report Number
6000001-2011-02814
Event Type
Malfunction
Date Received
April 13, 2011
Date of Event
June 3, 2010
Report Date
October 19, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 814:02 WAS CONFIRMED DURING PRODUCT EVALUATION IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY A DEFECTIVE PUMPHEAD MODULE. THE PUMPHEAD MODULE WAS REPLACED TO CORRECT THE REPORTED CONDITION. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.06.00. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

DURING REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER (B)(4), A COLLEAGUE INFUSION PUMP EXPERIENCED FAILURE CODE 814:02, WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE