FDA Adverse Event Malfunction Summary report: N

AED PLUS

MDR report key: 2052500 · Received March 15, 2011

Report

Report Number
1220908-2011-00662
Event Type
Malfunction
Date Received
March 15, 2011
Report Date
February 25, 2011
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K011541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AED PLUS DEFIBRILLATOR MKJ ZOLL MEDICAL CORPORATION AED PLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA