FDA Adverse Event Malfunction Summary report: N

POWERED WHEELCHAIR

MDR report key: 2052481 · Received March 15, 2011

Report

Report Number
1525712-2011-00093
Event Type
Malfunction
Date Received
March 15, 2011
Date of Event
February 1, 2009
Report Date
March 11, 2011
Manufacturer
INVACARE
Product Code
ITI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONSUMER OBSERVED SMOKE COMING FROM THE (B)(4) BOX. THE (B)(4) BOX WAS RECEIVED AND INSPECTED. NO EXTERNAL HEAT DAMAGE WAS OBSERVED ON THE BOX OR ITS WIRES. THE METAL BOX WAS DISASSEMBLED AND INTERNAL HEAT DAMAGE TO THE PC BOARD WAS OBSERVED. AS WITH ANY ELECTRONIC DEVICE, ELECTRONIC COMPONENT FAILURES OCCUR AS ONE DID IN THIS CONTROL BOX. HOWEVER FAILURES OF THIS TYPE ARE WELL CONTAINED INTERNALLY WITHIN THE METAL HOUSING OF THE CONTROL BOX. DURING THE EVENT THE CONTROL BOX LIKELY EMITTED SMOKE FOR A BRIEF TIME, HOWEVER NO FIRE OCCURRED AND NO HEAT DAMAGE WAS EXTERNAL TO THE COMPONENT. THE DEALER HAS REPLACED THE BOX, AND THE CHAIR REMAINS IN SERVICE.

Description of Event or Problem · 1

THE CONSUMER'S HUSBAND ALLEGES HE SAW SMOKE COMING FROM THE (B)(4) BOX. NO INJURY IS ALLEGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POWERED WHEELCHAIR 890.3860 ITI INVACARE TDXSP-CG

Patients

Seq Age Sex Outcome Treatment
1 68 YR