POWERED WHEELCHAIR
Report
- Report Number
- 1525712-2011-00093
- Event Type
- Malfunction
- Date Received
- March 15, 2011
- Date of Event
- February 1, 2009
- Report Date
- March 11, 2011
- Manufacturer
- INVACARE
- Product Code
- ITI
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
CONSUMER OBSERVED SMOKE COMING FROM THE (B)(4) BOX. THE (B)(4) BOX WAS RECEIVED AND INSPECTED. NO EXTERNAL HEAT DAMAGE WAS OBSERVED ON THE BOX OR ITS WIRES. THE METAL BOX WAS DISASSEMBLED AND INTERNAL HEAT DAMAGE TO THE PC BOARD WAS OBSERVED. AS WITH ANY ELECTRONIC DEVICE, ELECTRONIC COMPONENT FAILURES OCCUR AS ONE DID IN THIS CONTROL BOX. HOWEVER FAILURES OF THIS TYPE ARE WELL CONTAINED INTERNALLY WITHIN THE METAL HOUSING OF THE CONTROL BOX. DURING THE EVENT THE CONTROL BOX LIKELY EMITTED SMOKE FOR A BRIEF TIME, HOWEVER NO FIRE OCCURRED AND NO HEAT DAMAGE WAS EXTERNAL TO THE COMPONENT. THE DEALER HAS REPLACED THE BOX, AND THE CHAIR REMAINS IN SERVICE.
THE CONSUMER'S HUSBAND ALLEGES HE SAW SMOKE COMING FROM THE (B)(4) BOX. NO INJURY IS ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POWERED WHEELCHAIR | 890.3860 | ITI | INVACARE | TDXSP-CG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |